|  | HydroKnowz |  |
| HydroKnowz™ monitors use catalytic sensors to measure hydrogen in air up to 25,000 ppm (2.5% volume).
These sensors feature VOC filters, good long term stability, native linearity, increased tolerance of silicone-based contaminants, and are immune to overexposure damage.
Options include 3VDC (regulated) or 5-40VDC / 5-60VDC operation, dry relay contacts, alarm beeper, 0-3VDC / 4-20mA (proportional) or TTL (binary) output signal.
HydroKnowz™ Features:
- Catalytic sensor (with VOC filter)
- HCMOS microcontroller
- 0-25,000ppm hydrogen in air measurement
- 100ppm resolution
- 2 seconds start-up time
- 2 secs. T90 response time
- 5 secs. T10 recovery time
- Normal, Warning, Alarm & Error states
- Bi-color LED status indicator
- Optional NO or NC dry relay
- Manual reset or auto-resetting operation
- Optional 0-to-3V proportional output (8 bit)
- Optional 4-20mA proportional output (8 bit)
- Optional binary operation (TTL output)
- Optional alarm beeper
- PK-Port™ compatible digital communications
- User-settable alarm/warning via PK-Port™
- Digital FLASH-based calibration
- Each device factory pre-calibrated
- Field re-programmable firmware
- -40 to +80 Celsius / 0% to 99% R.H. operation
- Sensor & temperature error detection
- 1000 mW power consumption
- Optional 3VDC (regulated) input power version (Option "I1")
- Optional 5-40VDC input power version (Option "I2")
- Optional 5-60VDC input power version (Option "I3")
|
|  | |  |  |
|
|  | Specification Information |  |
HydroKnowz™ Details:
Warning & Alarm Levels· Alarm & warning PPMs specified at order time
· Alarm level from 2,000 to 20,000 ppm
· Warning level/action is optional
· Warning level from 20-80% of alarm level
· Levels are field-adjustable via PK-Port™
State Indications· Normal, warning, alarm & error state
· Bi-color LED indicates all states
· Optional beeper for alarm & error indication
· Errors-as-Alarms for fail safe operation
· Warning and/or alarm trigger relay/TTL
· Live monitoring & data logging via PK-Port
Reliability Features· Audio-visual or alarm-producing errors
· Sensor failure & disconnection detection
· Temperature out-of-bounds detection
· Digital calibration values delimited
· Low supply voltage inhibit
· Power fail detect via signal offset loss/relay
· Watchdog timer & software checksums
Environmental Requirements· Diffusion sampling
· 3 meters/second flow rate maximum
· 5% minimum atmospheric oxygen
· 0% to 99% R.H. (non-condensing) oper.
Calibration· All devices individually calibrated · FLASH-based calib. accessible via PK-Port · Periodic factory recertification avail. Incompatible Uses· Not for use in anaerobic environments
HydroKnowz™ Options:
PK-Port Interface· Access via interface connector pin
· PC (serial) interface module extra
· Real-time monitoring and data-logging
· Access/modify config. & cal. values
Mechanical· Open-frame circuit size: 61 x 48 x 26 mm
· Aluminum enclosure size: 74 x 82 x 26 mm
· Open-frame weight: 30 g
· Enclosed weight: 50 g
Error Handling Options
1. Audio-visual indication of errors [E1]
2. Errors trigger alarms (fail safe) [E2]
Output Signal Option
1. TTL - Norm.-high, Auto (self-clearing) [O1]
2. TTL - Norm.-low, Auto (self-clearing) [O2]
3. TTL - Norm.-high, Latch (Button reset) [O3]
4. TTL - Norm.-low, Latch (Button reset) [O4]
(Specify TTL signal to respond to warning state, alarm state, or both.)
5. Proportional output - 0 to 3VDC [O5]
0.5V safety offset (zero reading)
200 steps of 20mV each
Each step equals one resolution unit
6. Proportional output - 4-20mA [O6]
Audio Sounder Options1. No beeper [S0]
2. Included beeper [S1]
Relay Options
1. Relay not required [R0]
2. Normally-closed, Auto (self-clearing) [R1]
3. Normally-open, Auto (self-clearing) [R2]
4. Normally-closed, Latch (Button reset) [R3]
5. Normally-open, Latch (Button reset) [R4]
Specify relay to respond to warning state, alarm state, or both.
Enclosure Options
1. Open-frame (Circuit board only) [P0]
2. Anodized aluminum case (With flange) [P1]
Powering Options
1. 3 VDC (+/- 2%) [I1]
2. 5-to-40 VDC switch mode PSU [I2]
3. 5-to-60 VDC switch mode PSU [I3]
Interface Connector
1. Molex 6-pin Mini-Fit Jr. Header
- Pin 1: Input power (+)
- Pin 2: Ground (common)
- Pin 3: Output signal
- Pin 4: Dry relay contact #1
- Pin 5: Dry relay contact #2
- Pin 6: PK-Port interface
Custom connectors & harnesses are available with nominal NRE. Please call or email to discuss custom requirements. |
|  | |  |  |
|
|  | H-Tech Warranty & Guarantee |  |
H-Tech will work directly with you to address any manufacturer defects or other warranty issues in order to resolve any problem quickly and efficiently with minimum consumer effort.
H-Tech's Warranty Policy |
|  | |  |  |
|
|  | H-Tech Shipping Information |  |
• Estimated Ship Dimensions: 1" x 1" x 1" • Estimated Ship Weight: 1 lbs • Lead time: 14 days • Shipping origin: Vancouver, Canada Please contact us for pricing on this product. H-Tech's Shipping Policy |
|  | |  |  |
|
|  |